Potential of scan switch |
≤0.4 u V |
|
Resolution |
Max resolution of potential measuring 0.1 u V |
|
Max resolution of resistance measuring 0.1m Ω | |
|
Accuracy |
Accuracy of potential measuring≤0.01% |
|
Accuracy of resistance measuring≤0.01% | |
|
Accuracy of thermocouple inspection≤1.2℃ | |
|
| |
|
Accuracy of resistance inspection≤0.05℃ | |
|
Stability of constant temperature |
Temperature change after thermocouple inspection≤0.2℃/min |
|
Temperature change after thermoresistance inspection≤0.04℃/min | |
|
Cold automatic compensation |
When cold temperature(20±10)℃,error of compensation≤0.2℃ |
|
Inspection temperature |
Thermocouple 300~1100℃ |
|
Thermoresistance 0~300℃(include low-temperature thermocouple ) | |
|
Inspection count |
Thermocouple 1—5 branch same time |
|
Thermoresistance 1—5 branch same time | |
|
Once continuous inspection point |
Thermocouple 4 points |
|
Thermoresistance 3 points | |
|
Inspection time |
Thermocouple inspection, 30 min each 100℃ |
|
Working environment |
Environment temperature 20℃±3℃ |
|
Relative humidity ≤75% | |
|
Power supply |
AC220V±10% 50±1HZ |
|
Grounding resistance ≤4Ω |