Explosion-proof mark ExdIIBT4
Qualified number GYB05113
Use area 1area、2 area、IIA、IIB,T1~T4 dangerous area
Configuration features Through the PC, using the TMT181A configuration tool to configure
Interface RS232PC
Input Signal Type Thermal Resistance Pt100
range -200~650℃
Conversion Accuracy ≤0.08%
Output signal analog 4~20mA 20~4mA
Cold Junction Accuracy ±1℃(thermocouple)
Max load (V power -11.3V)/0.025A(current input)
Limit current ≤23mA
Response time 1s
Display digit four digits
Isolated voltage 2kVAC (input/output)
Power voltage 13.3~35VDC
Long-term stability ≤0.1℃/year,or 0.05%/year
Protection level IP66
Ambient temperature -20℃~+40℃