structure: Plate-mounted, portable-type
range:O~19.999mV-use thermocouple S(LB-3)
O~59.999mV-use thermocouple K (EU_2)
Set accuracy 0.3 grade
Regulating accuracy: ±O.1%
Amplifier drift: ≤4μV/24 hour
PID parameter:P:2~250% Continuously adjustable
I:O.5~10 minute Continuously adjustable
D:O~3 minute Continuously adjustable
Output signal:Pulse phase, Amplitude >3V width >30μs (20Ω load resistance)